Output power measurement of a blood pump considering pulsating flows
Sejong Chun and Jonghan Jin*
The Journal of Mechanical Science and Technology, vol. 26, no. 3, pp.793-798, 2012
Abstract : A blood flow simulator is one of the most useful tools for investigating the dynamic properties of a cardiovascular system. The blood
pump in a blood flow simulator generates pulsating flows by oscillating a membrane mechanically at different pulse rates. To evaluate
the performance of the blood pump properly, the flow characteristics of the pulsating flows should be considered. In this paper, two basic
indicators for evaluating the blood pump characteristics, the output power and the overall efficiency, were determined with consideration
of the pulsating flows, based on a phase-averaged analysis. These indicators were obtained by integrating closed contours in a P−Q
diagram. The measured output power was expected to be meaningful for understanding the dynamic effects caused by pulsating flows.
Keyword : Artificial blood pump; Blood flow simulator; Heart work; Oscillatory flow; Phase-averaging; Pulsating flow |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|