|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Structural analysis of a bileaflet mechanical heart valve prosthesis with curved leaflet Young Joo Kwon
The Journal of Mechanical Science and Technology, vol. 22, no. 11, pp.2038-2047, 2008
Abstract : Structural analysis, especially the thickness effect on the structural strength, of a bileaflet mechanical heart valve
prosthesis with curved leaflets is presented in this paper. Taking the wide variations of the blood flow pressure on the
leaflet surface, the structural stresses inside the leaflet and deflections of the leaflet are investigated by adopting both
linear and nonlinear structural analysis techniques for more accurate results comparison. The thickness of the curved
leaflet also varies considerably from 0.50 mm to 0.75 mm by 0.05 mm. These are very useful for the design of the mechanical
heart valve (MHV) prosthesis. Linear and nonlinear structural mechanic analyses for the leaflet of the MHV
prosthesis are conducted to predict the structural strength variation of the leaflet as the leaflet thickness changes. Analysis
results show that the structural strength of the leaflet decreases as the leaflet thickness becomes thinner and thinner,
and the nonlinear structural behaviors of thin leaflets are very conspicuous for large applied blood pressures. Hence,
these thin leaflets are not desirable for the in vivo use of the MHV prosthesis.
Keyword :
Bileaflet mechanical heart valve prosthesis; Linear and nonlinear structural analyses; Leaflet thickness; Blood pressure; Structural strength
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
JMST Editorial Office: #702 KSTC New Bldg, 22 7-gil, Teheran-ro, Gangnam-gu, Seoul 06130, Korea
TEL: +82-2-501-3605, E-mail: editorial@j-mst.org |
JMST Production Office: #702 KSTC New Bldg, 22 7-gil, Teheran-ro, Gangnam-gu, Seoul 06130, Korea
TEL: +82-2-501-6056, FAX: +82-2-501-3649, E-mail: editorial@j-mst.org |
|
|
|
|